Locking your nmr sample this handout contains instructions on how to lock on any deuterated solvent.
Nmr lock roof.
Manual locking in vnmr is controlled via acqi window.
The most convenient way for the standard user to lock is semi automatically using the xwin nmr command lock.
Tune z0 if needed.
If it still does not lock try using the command ii to reset communications between console and computer.
Now click lock button.
Given an nmr spectrum with two coupled nuclei one can often observe so called roofing an asymmetry in the multiplet.
The lock on a modern varian spectrometer is essentially a separate lock sweep spectrometer that is specific to deuterium.
Click and hold in 1 mode.
One of the most important factors to get a good quality nmr spectrum is shimming.
Turn the lock off by typing lock off then turn it back on by typing lock 3.
Nrm 1750 panels snap lock floating clip system.
Performance clip for ul 90 rated roof installations.
This spectrometer is typically tuned to the deuterium nmr resonance frequency.
Fasteners for roofing applications use wood type screws fastened on 36 centers.
This opens a new window in which the lock trace now appears.
The field lock is a separate nmr spectrometer within your spectrometer.
Reload the default shim file by typing rsh default 4.
Lock on the solvent 5.
The window will probably change into something like you can see on figure 2 or 3.
Assuming that lock is not yet well tuned.
If the signal gets noisy reduce lock gain instead of power.
For example here is a 300 mhz proton spectrum of 1 bromo 2 methoxyethane where the inset shows the triplets corresponding to the adjacent ce ch2 groups.
To get there type acqi in the command line or click acqi button a window as shown on fig.
To open the lock display type lockdisp.
Other screw types are available applications over structural steel and steel decking.
Once the lock is set turn the power and gain back to their previous values.
If the lock system didn t engage and the green light is blinking try increasing temporarily the power or the gain.
To lock the lock signal should be a reasonable size centered in the lock display window show significant ringing and be phased so the largest part of the signal is positive.
When unlocked the lock display shows the lock signal as the field is swept back and forth.
To display the lock signal enter lockdisp.
Locking is simply the adjustment of the deuterium lock field z0.
Before running an nmr experiment it is also necessary to lock and shim the magnetic field.